48 Volara™ System, Model PVL1 User Manual (196286 REV C)
Device Settings
ENABLE OR DISABLE CLINICAL ACCESS
Clinical Access allows facility-authorized persons to access and configure
these advanced features:
• Adjustment of the maximum pressure ceiling setting available to the
user.
• Modification of the preset therapy settings.
• Import of the device settings
• Set up and pairing of Bluetooth® devices, such as a barcode reader or
pulse oximeter
• Upgrade of the system firmware (See the service manual for details.).
NOTE:
Import of the Health Level 7 (HL7) standard file
1. Press Device, and press Controls.
2. Slide the setting for Clinical Access to
ON.
• To disable, slide the setting to OFF.
The user will not be able to access
the manual modes or modify the
preset therapy plans.
3. Enter the key code, and press Enter. If
applicable, contact your facility
administrator to get the key code.
4. When complete, press the Home menu
control to exit.
196654__C_2019-Apr-29_Cleaned