86 Volara™ System, Model PVL1 User Manual (196286 REV C)
Maintenance
NEBULIZER FILTER
Examine the nebulizer filter every month, and clean it as necessary.
Replace the inlet filter every twelve (12) months.
To clean the filter, do as follows:
1. Power off the control unit.
2. Remove the inlet filter from the back of the control unit.
NOTE:
If the filter is damaged, replace it. See “Replacement Parts/Kits” on
page 78.
3. Wash the filter with mild
detergent and warm water,
and then rinse it with clean
water.
4. Allow the filter to dry
completely, and then install it
into the control unit.
FUSE
If a fuse needs to be replaced, do as follows:
196654__C_2019-Apr-29_Cleaned