EasyManuals Logo

Hill-Rom Volara PVL1 User Manual

Hill-Rom Volara PVL1
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #99 background imageLoading...
Page #99 background image
Volara™ System, Model PVL1 User Manual (196286 REV C) 91
Troubleshooting
Caution Messages
Caution messages show on-screen to
let you know that an issue has
occurred. Follow the on-screen
instructions to troubleshoot the issue.
If an error code shows, see “Solutions
for Cautious Messageson page 91.
Solutions for Cautious Messages
Error Code Do this:
007, 013, 020 1. Press return after you read the on-screen
message.
2. Power off the control unit, and then power
it on.
011 1. Make sure the power cord is fully plugged
into the electrical inlet on the back of the
control unit. If necessary, disconnect the
power cord from the control unit, and then
connect it.
2. Make sure the power cord is securely
plugged into an AC power.
3. Press return after you read the on-screen
message.
014 1. Make sure the correct patient circuit is set
up for the required therapy. For details, see
Assemble and Connect the Patient Circuit
on page 16.
2. Press return after you read the on-screen
message.
015, 016, 017, 018 1. Press return after you read the on-screen
message.
2. Power off the control unit and let it cool
down for at least 5 minutes. Power on the
control unit.
3. If the same message shows, power off the
control unit and let it cool down for at least
30 minutes. Power on the control unit.
196654__C_2019-Apr-29_Cleaned

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hill-Rom Volara PVL1 and is the answer not in the manual?

Hill-Rom Volara PVL1 Specifications

General IconGeneral
BrandHill-Rom
ModelVolara PVL1
CategoryMedical Equipment
LanguageEnglish

Related product manuals