102 Volara™ System, Model PVL1 User Manual (196286 REV C)
Specifications
ELECTROMAGNETIC COMPATIBILITY GUIDANCE
The Volara™ System, Model PVL1 is suitable for the electromagnetic
environment settings specified in the tables that follow.
WARNING:
Warning—The Volara™ System, Model PVL1 should not be used adjacent
to or stacked with other equipment. If adjacent or stacked use is
necessary, the Volara™ System should be observed to verify normal
operation. If operation is not normal, the Volara™ System or any other
equipment should be moved.
Use of accessories, transducers, and cables, other than those specified or
provided by the manufacturer of this equipment, could result in increased
electromagnetic emissions or decreased electromagnetic immunity of
this equipment and result in improper operation.
Guidance and Manufacturer's Declaration—Electromagnetic
Emissions
The Volara™ System, Model PVL1 is intended for use in the electromagnetic environment
specified below. The customer or user of the Volara™ System, Model PVL1 should make
sure it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment—Guidance
RF Emissions
CISPR 11
Group 1 The Volara™ System uses RF energy only for its
internal functions. Therefore, its RF emissions
are low and are not likely to cause any interfer-
ence in nearby electronic equipment.
RF Emissions
CISPR 11
Class B The Volara™ System is suitable for use in all
establishments, including domestic establish-
ments, and those directly connected to the pub-
lic low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Complies
196654__C_2019-Apr-29_Cleaned