90 Volara™ System, Model PVL1 User Manual (196286 REV C)
Troubleshooting
Warning Messages
Warning messages show on-screen
to let you know that a critical issue
has occurred. Follow the on-screen
instructions to troubleshoot the
issue. If an error code shows, see
“Solutions for Warning Messages” on
page 90.
Solutions for Warning Messages
Error Code Do this:
001, 002, 003, 004,
005, 006, 008, 032
Power off the control unit, then power it on
again.
009
1. Make sure only the Hill-Rom approved
battery (194566S) is used.
2. Press return after reading the on-screen
message.
3. Make sure the battery is correctly installed in
the control unit. If necessary, remove the
battery and install it. Remove the battery and
install it. See “Install the Battery” on page 67“.
010 1. Power off the control unit.
2. Make sure the nebulizer connected to the
system is functioning properly. Change to a
new nebulizer machine, if required.
3. Power on the control unit.
4. Start the therapy again.