Section
3: Functions and features Series 2600B System SourceMeter® Instrument
Reference Manual
3-30 2600BS-901-01 Rev. B / May 2013
Source and measurement delays
Whenever the source-measure unit (SMU) outputs a source value in a sweep, it also applies the
programmed source delay. The default source delay is zero (0) seconds. Set an additional source
delay using the smuX.source.delay attribute.
Whenever the SMU performs a measurement in a sweep, it also applies any configured
measurement delays. Use the smuX.measure.delay attribute to program a specific measurement
delay. The default measurement delay varies by measure range.
Initiating and running sweeps
To run a sweep, you must configure the number of sweep points to output and the number of sweeps
to perform. Use the trigger count to set the number of sweep points to output. Use the arm count to
set the number of times to perform the sweep. See Triggering (on page 3-32
) for more information.
Examples:
To start a sweep, use the smuX.trigger.initiate() function. Sweeps are overlapped
operations, so you can use the waitcomplete() function as a way to suspend further operation
until the sweep is complete.
To sweep 15 source points:
smua.trigger.count = 15
To perform eight sweeps:
smua.trigger.arm.count = 8
Aborting a sweep
The smuX.abort() function can be used to terminate all overlapped operations on a
source-measure unit (SMU), including sweeps. It returns the SMU to the idle state of the remote
trigger model. See Triggering (on page 3-32
) for more information.
Sweeping using factory scripts
Factory script functions that perform linear staircase, logarithmic staircase, and list sweeps are listed
in Introduction to TSP operation (on page 5-1
). You can use the factory script functions to perform
and execute simple sweeps, or use them as examples for programming your own custom sweeps.
Front panel
To run a sweep from the front panel:
1. Press the LOAD key, and then select FACTORY.
2. Select the name of the test to run.
3. Press the RUN key, and then follow the display prompts to complete the test.
See Introduction to TSP operation (on page 5-1) for more information about using factory scripts.
Press the RECALL key to access sweep data stored in dedicated reading buffer 1. See Reading
buffers (on page 3-6) for more details about the buffer.

Table of Contents

Other manuals for Keithley 2614B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2614B and is the answer not in the manual?

Keithley 2614B Specifications

General IconGeneral
BrandKeithley
Model2614B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals