Series 2600B
System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. B / May 2013 7-197
smuX.makebuffer()
This function creates a reading buffer.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
bufferVar = smuX.makebuffer(bufferSize)
bufferVar
The created reading buffer
X
Source-measure unit (SMU) channel (for example,
smua.makebuffer()
applies
to SMU channel A)
bufferSize
Maximum number of readings that can be stored
Details
Reading buffers can be created and allocated dynamically using this function. Use bufferSize to designate the
number of readings the buffer can store.
Dynamically allocated reading buffers can be used interchangeably with the smuX.nvbufferY buffers.
A reading buffer can be deleted by setting all references to the reading buffer equal to nil, then running the
garbage collector (see the collectgarbage() function in Standard libraries (on page 6-26
)).
Example
mybuffer2 = smua.makebuffer(200)
Creates a 200 element reading buffer (mybuffer2)for
SMU channel A.
Also see
collectgarbage() in Base library functions (on page 6-26)
Remote reading buffer programming (on page 3-11)
savebuffer() (on page 7-161)
smuX.nvbufferY (on page 7-215)

Table of Contents

Other manuals for Keithley 2614B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2614B and is the answer not in the manual?

Keithley 2614B Specifications

General IconGeneral
BrandKeithley
Model2614B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals