7: TSP command reference Series 2600B System SourceMeter® Instrument
7-104 2600BS-901-01 Rev. B / May 2013
Details
This function only initiates configured pulse trains assigned to a valid tag. Configure the pulse before initiating it
using one of the ConfigurePulse* functions (refer to the Also see section).
Example
smua.source.rangev = 5
smua.source.rangei = 1
smua.source.levelv = 0
smua.measure.rangev = 5
smua.measure.rangei = 1
smua.measure.nplc = 0.01
smua.measure.autozero = smua.AUTOZERO_ONCE
smua.nvbuffer1.clear()
smua.nvbuffer1.appendmode = 1
smua.source.output = smua.OUTPUT_ON
f1, msg1 = ConfigPulseVMeasureI(smua, 0, 5,
1, 0.002, 0.2, 10, smua.nvbuffer1, 1)
if f1 == true then
f2, msg2 = InitiatePulseTest(1)
print("Initiate message:", msg2)
else
print("Config errors:", msg1)
Configure System SourceMeter
instrument channel A to generate
a pulse train. If no errors are
encountered, initiate the pulse
train. Channel A pulses voltage
from a bias level of 0 V to a pulse
level of 5 V. The pulse level is
present for 2
for 200 ms, with a 1
A total of 10 pulses is generated,
and the measurement data is
stored in smua.nvbuffer1. This
pulse train is assigned to tag =
1.
Also see
ConfigPulseIMeasureV() (on page 7-33)
ConfigPulseVMeasureI() (on page 7-39)
ConfigPulseIMeasureVSweepLin() (on page 7-35)
ConfigPulseVMeasureISweepLin() (on page 7-41)
ConfigPulseIMeasureVSweepLog() (on page 7-37)
ConfigPulseVMeasureISweepLog() (on page 7-43)
KIPulse factory script (on page 5-21)
InitiatePulseTestDual()
This KIPulse factory script (on page 5-21) function configures initiates the pulse configuration assigned tag1 and
tag2.
Type TSP-Link accessible
Affected by Where saved Default value