Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-30 2600BS-901-01 Rev. B / May 2013
Example 2
printbuffer(1, 6,
smua.nvbuffer1.sourcevalues)
Print the sourced value of the first 6 readings
stored in source-measure unit (SMU) A,
buffer 1.
Example output:
1.00000e-04, 1.00000e-04,
1.00000e-04, 1.00000e-04,
1.00000e-04, 1.00000e-04
Also see
bufferVar.measurefunctions (on page 7-22)
bufferVar.measureranges (on page 7-23)
bufferVar.n (on page 7-24)
bufferVar.readings (on page 7-25)
bufferVar.sourcefunctions (on page 7-26)
bufferVar.sourceoutputstates (on page 7-27)
bufferVar.sourceranges (on page 7-28)
bufferVar.statuses (on page 7-30)
bufferVar.timestamps (on page 7-32)
Reading buffers (on page 3-6)
bufferVar.statuses
This attribute contains the status values of readings in the reading buffer.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R)
Yes
Clearing the buffer
See Details
Not applicable
Usage
statusInformation = bufferVar.statuses[N]
statusInformation
The status value when reading N of the specified buffer was acquired
bufferVar
The reading buffer; can be a dynamically allocated user-defined buffer or a
dedicated reading buffer
N
The reading number (1 to
bufferVar
.n)

Table of Contents

Other manuals for Keithley 2614B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2614B and is the answer not in the manual?

Keithley 2614B Specifications

General IconGeneral
BrandKeithley
Model2614B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals