7: TSP command reference Series 2600B System SourceMeter® Instrument
7-30 2600BS-901-01 Rev. B / May 2013
Example 2
smua.nvbuffer1.sourcevalues)
Print the sourced value of the first 6 readings
stored in source-measure unit (SMU) A,
buffer 1.
Example output:
1.00000e-04, 1.00000e-04,
1.00000e-04, 1.00000e-04,
Also see
bufferVar.measurefunctions (on page 7-22)
bufferVar.measureranges (on page 7-23)
bufferVar.n (on page 7-24)
bufferVar.readings (on page 7-25)
bufferVar.sourcefunctions (on page 7-26)
bufferVar.sourceoutputstates (on page 7-27)
bufferVar.sourceranges (on page 7-28)
bufferVar.statuses (on page 7-30)
bufferVar.timestamps (on page 7-32)
Reading buffers (on page 3-6)
bufferVar.statuses
This attribute contains the status values of readings in the reading buffer.
Type TSP-Link accessible Affected by Where saved Default value
Usage
statusInformation = bufferVar.statuses[N]
The status value when reading N of the specified buffer was acquired
The reading buffer; can be a dynamically allocated user-defined buffer or a
dedicated reading buffer
bufferVar