7: TSP command reference Series 2600B System SourceMeter® Instrument
7-364 2600BS-901-01 Rev. B / May 2013
tsplink.trigger[N].wait()
This function waits for a trigger. This function is not available on the Models 2604B/2614B/2634B.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
triggered = tsplink.trigger[N].wait(timeout)
Trigger detection indication; set to one of the following values:
true: A trigger is detected during the timeout period
: A trigger is not detected during the timeout period
The trigger line (1 to 3)
The timeout value in seconds
Details
This function waits up to the timeout value for an input trigger. If one or more trigger events were detected since
the last time tsplink.trigger[N].wait() or tsplink.trigger[N].clear() was called, this function
returns immediately.
After waiting for a trigger with this function, the event detector is automatically reset and rearmed. This is true
regardless of the number of events detected.
Example
triggered = tsplink.trigger[3].wait(10)
print(triggered)
Waits up to 10 seconds for a trigger
on TSP-Link
®
line 3.
If false is returned, no trigger was
detected during the 10-second
timeout.
If true is returned, a trigger was
Also see
tsplink.trigger[N].clear() (on page 7-356)
tsplink.writebit()
This function sets a TSP-Link synchronization line high or low. This function is not available on the Models
2604B/2614B/2634B.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
tsplink.writebit(N, data)
The trigger line (1 to 3)
The value to write to the bit:
• Low: 0
•