7: TSP command reference Series 2600B System SourceMeter® Instrument
7-352 2600BS-901-01 Rev. B / May 2013
tsplink.master
This attribute reads the node number assigned to the master node. This attribute is not available on the Models
2604B/2614B/2634B.
Type TSP-Link accessible Affected by Where saved Default value
Usage
masterNodeNumber = tsplink.master
The node number of the master node
Details
After doing a TSP-Link reset (tsplink.reset()), use this attribute to access the node number of the master in
a set of instruments connected over TSP-Link.
Example
LinkMaster = tsplink.master
Store the TSP-Link master
node number in a variable
Also see
tsplink.reset() (on page 7-354)
tsplink.node
This attribute defines the node number. This attribute is not available on the Models 2604B/2614B/2634B.
Type TSP-Link accessible Affected by Where saved Default value
Usage
nodeNumber = tsplink.node
tsplink.node = nodeNumber
The node number of the instrument or enclosure (1 to 64)
Details
This attribute sets the TSP-Link node number and saves the value in nonvolatile memory.
Changes to the node number do not take effect until tsplink.reset() from an earlier TSP-Link instrument is
executed on any node in the system.
Each node connected to the TSP-Link system must be assigned a different node number.
Example
Sets the TSP-Link node for this instrument to
number 3.
Also see
tsplink.reset() (on page 7-354) tsplink.reset() (on page 7-354)
tsplink.state (on page 7-355)