Page #627 background imageLoading...
Page #627 background image
Series 2600B
System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. B / May 2013 7-287
In addition to the above constants, operationRegister can be set to the numeric equivalent of the bit to set.
For example, to set bit B10, set operationRegister to 1024.
Example
status.operation.instrument.trigger_blender.enable = 1024
Uses a decimal value to set
the TRGOVR bit of the
operation status trigger
blender summary enable.
Also see
Operation Status Registers (on page E-9)
status.operation.instrument.trigger_blender.trigger_overrun.* (on page 7-287)
status.operation.instrument.trigger_blender.trigger_overrun.*
This attribute contains the operation status trigger blender overrun register set.
Type TSP-Link accessible Affected by Where saved Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW) Yes Status reset Not saved 0
.ptr (RW)
Yes
Status reset
Not saved
126 (All bits set)
Usage
operationRegister =
status.operation.instrument.trigger_blender.trigger_overrun.condition
operationRegister =
status.operation.instrument.trigger_blender.trigger_overrun.enable
operationRegister =
status.operation.instrument.trigger_blender.trigger_overrun.event
operationRegister =
status.operation.instrument.trigger_blender.trigger_overrun.ntr
operationRegister =
status.operation.instrument.trigger_blender.trigger_overrun.ptr
status.operation.instrument.trigger_blender.trigger_overrun.enable =
operationRegister
status.operation.instrument.trigger_blender.trigger_overrun.ntr =
operationRegister
status.operation.instrument.trigger_blender.trigger_overrun.ptr =
operationRegister
operationRegister
The status of the operation status trigger blender overrun register; a zero (0)
indicates no bits set (also send 0 to clear all bits); other values indicate various bit
settings

Table of Contents

Other manuals for Keithley 2614B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2614B and is the answer not in the manual?

Keithley 2614B Specifications

General IconGeneral
BrandKeithley
Model2614B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals