Series 2600B
System SourceMeter® Instrument Reference Manual Section 7:
TSP command reference
2600BS-901-01 Rev. B / May 2013 7-219
Details
This function generates and activates new calibration constants for the given range.
The positive and negative polarities of the source must be calibrated separately. Use a positive value for range to
calibrate the positive polarity and a negative value for range to calibrate the negative polarity. Do not use 0.0 for
a negative calibration point; 0.0 is considered to be a positive number.
Typically, the two calibration points used will be near zero for calibration point 1 and 90% of full scale for
calibration point 2. Full scale for calibration point 2 should be avoided if the source of the SMU is substantially
out of calibration.
The two reference measurements must be made with the source using the active calibration set. For example,
source a value, measure it, and do not change the active calibration set before issuing this command.
The new calibration constants are activated immediately but they are not written to nonvolatile memory. Use
smuX.cal.save() to save the new constants to nonvolatile memory.
The active calibration constants stay in effect until the instrument is power cycled or a calibration set is loaded
from nonvolatile memory with the smuX.cal.restore() function.
This function is disabled until a successful call to smuX.cal.unlock() is made.
Example
smua.source.calibratev(1, 1e-10, 1e-5, 0.9,
0.903)
SMU channel A calibrates the voltage
source using the following values:
calibrate the 1 V range, 1e-10 for +zero
source output value, 1e-5 for +zero DMM
measurement reading, 0.9 for +FS source
output value, and 0.903 for the +FS DMM
measurement reading.
Also see
Calibration (on page B-1)
smuX.cal.restore() (on page 7-188)
smuX.cal.save() (on page 7-189)
smuX.cal.unlock() (on page 7-190)
smuX.measure.calibrateY() (on page 7-200)
smuX.source.compliance
This attribute contains the state of source compliance.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R) Yes Not applicable Not saved Not applicable
Usage
compliance = smuX.source.compliance
compliance
The state of source compliance
X
Source-measure unit (SMU) channel (for example, smua.source.compliance
applies to SMU channel A)

Table of Contents

Other manuals for Keithley 2614B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2614B and is the answer not in the manual?

Keithley 2614B Specifications

General IconGeneral
BrandKeithley
Model2614B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals