Display & Cine Review 6-7
to switch and display the images in either memory, so you can compare images in
Split mode.
The memory capacity is evenly distributed as per splitting number, as shown in the following
table (Taking B images in low density as an example, the capacity of B cine memory is N
frames):
Imaging
Mode
Split
Single-B/ Color Dual Quad
Auto
One memory, with
capacity N frames.
The memory splits
into two, with
capacity N/2 frames
each
The memory splits
into four, with
capacity N/4 frames
each
Split
The memory splits
into two, with
capacity N/2
frames each
The memory splits
into two, with
capacity N/2 frames
each
The memory splits
into four, with
capacity N/4 frames
each
6.4.2 Cine Memory Clear
In the following conditions, the cine review memory will be cleared:
Start an exam of a new patient.
Start a new exam for the same patient.
Switching the probe (if the cine memory is split, only the cine memory corresponding
to the currently activated window is cleared)
Changing the exam condition (if the cine memory is split, only the cine memory
corresponding to the currently activated window is cleared)
Imaging Mode Switching: include imaging mode switching among B, M, color, PW,
display mode switching etc.
Parameters modification, including:
Parameters that can result in imaging region or direction changing, such as
depth, FOV, trapezoid, steer, zoom and so on.
Parameters that can result in image frame changing, such as line density, focus
number and so on.
Change the scanning speed
2D image cleared
Unfreeze the image: after the image is unfrozen, images stored in the cine memory
will be cleared, but if the cine memory is split, only the currently cine memory
corresponding to the activated window will be cleared.
Open/ close the image file that occupies cine memory.
6.5 Preset
Open [Setup] →[System Preset]→[General] to preset the cine storage length.
For time without ECG: 1-60s
For time with ECG loop,the unit is beat: 1-16