DICOM 11-3
11.1.2 DICOM Preset
To set the DICOM service and server properties.
1. Press <Setup> to enter the [Setup] menu.
2. Select [DICOM Preset].
3. Preset local DICOM properties and DICOM server.
Localhost DICOM Service Property
Name
Description
AE Title
Application entity title of the ultrasound system.
The AE title here should be the same with the one of the acceptable
SCU set in the server.
Port
DICOM communication port, which should be the same with the one
in the server.
PDU
Maximum PDU data package size ranging from 16384 to 65536.
The default value is 32768.
Enter AE Title, Port, PDU according to the actual situation.
Note:
AE Title should be the same with the SCU AE Title preset in the server (PACS/ RIS/
HIS), for example, if the AE Title of the server preset in the print server is AAA, and