Probes and Biopsy 13-27
<1>
bracket
Used for installing the needle-guided bracket on the
transducer.
<2>
Groove of the needle-
guided bracket
Matches with the tab of the transducer.
There are 3 types of angles available to be adjusted.
<4>
Angle shift sign (30°, 40°,
50°)
Matches with the biopsy angle
(30°, 40°, 50°).
<5> Angle pinch nut Used for fixing the angle lock at a chosen angle.
<6> Angle block
Determines the angle of the biopsy; different
specifications of blocks can be used.
<7> Guiding block
Used for installing the needles of different specifications,
5 types of needles are available.
<8>
Specification of guiding
block (14G)
Matched with the corresponding biopsy needle (14G).
<9>
needle
Used for installing the needles.
<10>
Pinch nut of needle-
guided bracket
Used for locking the needle-guided bracket and the
transducer.
NGB-022
13.2.3 Needle-guided Bracket Inspection and Installation
Inspection of the Needle-guided Bracket
Be sure to perform inspections before and after use of the needle-guided bracket. If an
abnormality is found on the needle-guided bracket, immediately stop using it and contact
MINDRAY Customer Service Department or sales representative.
1. Sterilize the needle-guided bracket before and after use.
2. Confirm that the needle-guided bracket is free of damage, deformation, stripping,
malfunction, loose or missing parts.