11-2 DICOM
11.1 DICOM Preset
11.1.1 Network Preset
NOTE
DO Not edit IP address in Network Preset page as the network task is executed,
otherwise it fails. View if there is any undergoing task in task manager of the system.
Preset local TCP/IP settings.
1. Press <Setup> to enter the [Setup] menu.
2. Select [Network Preset].
3. Local TCP/ IP preset items are described as follows:
Name
Description
Current Net Adapter
To select network connection mode.
DHCP
/ Static
DHCP: IP address will be automatically obtained from DNS server;
Static: you need to enter the IP address.
IP Address
IP address of the system.
Subnet Mask Used to set different network segment.
Gateway
Used to set the gateway IP.
Note:
The IP address of the system should not be occupied by other devices in the network,
or the DICOM function will fail.
After you changed “Station Name”, you need to restart the system to make the setting
effective (press the power button and select “Shut Down” and restart the system after
a while).