General information
3.3 Measuring high oxygen concentrations
16
Quick Start
Compact Operating Instructions, 05/2018, A5E31805656-06
Measuring high oxygen concentrations
Note
Measuring high oxygen concentrations
For the measurement of high O
2
concentrations in excess of 25%, it is absolutely necessary
that all parts that come in contact with the sample gas are free of oil and grease and do not
contain any particles that could react with O
2
.
We highly recommend you use specially cleaned analyzer modules with the "Cleaned for O
2
r such measurements.