Commissioning
7.2 General information on commissioning
74
Quick Start
Compact Operating Instructions, 05/2018, A5E31805656-06
Hot surfaces resulting from hot sample gases and heated devices
Danger of burns resulting from surface temperatures above 70 °C (155 °F).
• Take appropriate protective measures, for example contact protection.
• Make sure that the maximum ambient temperature is not exceeded by implementi
ng
protective measures. Refer to the information in the section "Technical specifications
(Page 119)".
Formation of condensation in the device
Condensation can be caused by a high difference in temperature between transportation
and storage and the installation location.
• Before taking the device into operation let the device adapt for several hours in the new
environment.
General information on commissioning
Note
Ambient temperature
Ensure that the ambient temperature is within the permissible range during operation. The
corresponding information is available in the section "
Technical specifications (Page 119)".
The technical specificati
ons of the device can deviate outside these temperature limits.
Operability of the device is no longer ensured at deviations of more than 5
K.
Note
Vibrations at the installation site disturb the measured signal
When you use OXYMAT
7 and/or ULTRAMAT 7 analyzer modules, vibrations, especially in
the frequency range from 6
Hz to 15 Hz, can influence the measured signal.
Select an installation location which is vibration-free.
If necessary, change the measuring frequency settings:
– Make sure that the measuring frequencies do not match the fundamental of the
installation site nor one of its harmonic multiples.
– You must also ensure that the measuring frequency differs from half the basic
frequency of the vibration.