7.3.3 Selecting the calibration method
Figure 7-3 Quick start step 3
The module can always be calibrated in two different ways:
● Using reference weights: in the case of a calibration with weights, mechanical influences of
the scale construction are also partially taken into account.
● Without weights, using the technical specifications of the connected load cell(s): in the case
of automatic calibration, the accuracy of the scale is influenced by the mechanical properties
to a greater extent than with calibration using reference weights.
With both methods, make sure that the mechanical properties of the scale are flawless prior to
calibration.
Commissioning
7.3 5-minute quick start with the operator panel and the Ready-for-Use software
SIWAREX WP321
40 Operating Instructions, 08/2019, A5E33715669A-AD