Variable Note Type Lengt
h
(byte
s)
Rw Default Min. Max.
Maximum
positive zero‐
ing limit
(Page 64)
Positive range of semi-automatic
zeroing device [in % of max. weighing
range]
REAL 4 rw 3.0 0 100.0
Standstill
range
(Page 64)
Standstill range (in d) REAL 4 rw 0.1 0 maximum
number
range+
Standstill
time
(Page 65)
Standstill time 1 in ms TIME 4 rw 2 000 10 10 000
Standstill
waiting time
(Page 65)
Waiting time until standstill.
0: standstill-dependent scale com‐
mand, if there is no standstill, imme‐
diately rejected.
> 0: Maximum waiting time until com‐
mand is executed
TIME 4 rw 2 000 0 10 000
Limit frequen‐
cy low pass
filter 1
(Page 66)
Low-pass filter cut-off frequency:
0: Filter disabled
REAL 4 rw 1 0.01 200
Order no. low
pass filter 1
(Page 66)
Filter number UINT 2 rw 4 0 4
Reserve Reserve UINT 2 rw 0 - -
Limit frequen‐
cy low pass
filter 2
(Page 66)
Low-pass filter limit frequency:
0: Filter disabled
REAL 4 rw 0 0.01 200
Order no. low
pass filter 2
(Page 66)
Filter number UINT 2 rw 4 0 4
Mean value
filter depth
(Page 67)
Filter for digit values,
Permissible filter depth: 0 … 250
UINT 2 rw 10 0 500
1)
Parameter for calculation of calibration points
8.4.1 Scale name
You can select any name, but it may not exceed 12 characters. You can enter any designation.
Scale parameters and functions
8.4 DR 3 calibration parameters
SIWAREX WP321
62 Operating Instructions, 08/2019, A5E33715669A-AD