7.3.5 Setting calibration points
Figure 7-5 Quick start step 5a
Carry out the calibration commands at the end of the quick start:
1. Carry out the "Set calibration weight 0" command. The "Calibration weight 0" defined in step
4 is now visible in the display.
2. Place the "Calibration weight 1" defined in step 4 on the scale construction, and execute the
"Set calibration weight 1" command.
3. If an "Calibration weight 2" was selected:
Place "Calibration weight 2" defined in step 4 on the scale construction, and execute the
"Set calibration weight 2" command.
4. Calibration of the scale is now complete. Return to the start screen by clicking on the house
icon.