at the lower rear panel). To avoid loss of image quality, it is compul-
sory to connect this point to ground (earth).
• Potentially hazardous electrical voltage exists inside the equipment.
Unauthorized personnel DONOT attempt to open the cover of the
equipment.
• The equipment must not be used in the presence of inflammable gases
(e.g. anesthetic gases and nitrous oxide) to avoid explosion hazard.
• Devices must be powered from a receptacle marked "hospital grade"
before being connected to the system directly. In case "hospital grade"
receptacles are not available, use isolation transformers instead.
Warning!
Electrostatic discharge (ESD) may cause electric shock or
damage the equipment. Observe the following precautions:
1. Prior to repairing or cleaning the equipment, ensure
that the system has been turned off and the power ca-
ble unplugged.
2. The equipment must be grounded correctly during op-
eration. Use anti-ESD spray on the ground if possible.
• The leakage current of the entire system including all auxiliary equip-
ments must not exceed the limit as stated in IEC 60601-1.
• The equipment may interfere with or be interfered by other high fre-
quency devices (e.g. medical lasers). Extra safety measures must be
taken if other HF devices have to be used nearby.
• Use the couplant shipped with the equipment or any other couplant
recommended by SonoScape. Use any unrecommended couplant
may damage the probe and void the warranty.
2.4 Mechanical Safety
1. Prior to using the equipment, place it horizontally and lock the wheels.
2. Take care when moving the equipment. Failure to follow the pre-
cautions listed below could result in injury, uncontrolled motion and
costly damage.
∗ Take the following precautions before transporting the system:
1) Ensure that the system is powered off and with power cable un-
plugged.
P/N: 4720-0034-01A
2-3