3. Check the box in front of "Set fix IP address".
4. Set "Static IP" to 192.168.254.182.
Set "Net mask" to 255.255.255.0.
Set "Gateway" to 192.168.254.254.
Set "Printer IP" to 192.168.254.183.
Keep other items unchanged.
5. Select "Save Setting" to save settings and exit.
9.3.2.4 Post Installation
1. Connect the printer to a SonoScape ultrasound system, whose IP ad-
dress has been configured correctly, with an Ethernet cable.
2. Test the print function on the ultrasound system.
3. If the system passed the test, put a green label on it.
4. Otherwise review the IP setting of the ultrasound system, and/or re-
install the printer driver.
9.3.3 Network Connection with PC
Network connection between the S6 system and a PC can be achieved by
either direct connection or indirect connection through router.