12.2 Complete Wash Cycle Program
The
Complete Wash Cycle
program is validated to clean, with certainty, up to 12.0 kg
(26.4lb) of instruments using the Unity, and up to 16.0 kg (35.2 lb) of instruments using the
Unity20.
Program number P1
Program name Complete Wash Cycle
Supervisor editable no (locked)
Stage Parameter Factory setting
Pre-wash
Duration 1 minutes
Times performed 1 time
Degassing Duration
1 minute (Unity)
4 minutes (Unity 20)
Washing
Chemical dose
100 ml (3.38 fl.oz) Unity
45 ml (1.52 fl.oz) Unity 20
Temperature +40 °C (+104 °F)
Duration
1 minute (Unity)
6 minutes (Unity 20)
Rinse
Temperature 0 °C (+32 °F)
1
Duration 1 minutes
Final Rinse
Temperature 0 °C (+32 °F)
1
Duration 1 minutes
1
A setting of 0 °C (+32 °F) is interpreted at 'temperature not important' by the equipment. The
tank will fill with water from the supply hoses but the equipment will not attempt to regulate the
incoming water temperature, or adjust the temperature of the water in the tank.
WASH PROGRAMS 103
InnoWave Unity & Unity 20 Operator Manual

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Steris InnoWave Unity and is the answer not in the manual?

Steris InnoWave Unity Specifications

General IconGeneral
BrandSteris
ModelInnoWave Unity
CategoryMedical Equipment
LanguageEnglish

Related product manuals