12.5 Decontamination Program
The factory-set Decontamination program can be used to perform a full decontamination of
equipment, following the procedure described in
Performing a Full Decontamination
on
page71 . This wash program is locked and can only be adjusted by a manufacturer-
approved service technician.
Program number P6
Program name Decontamination
Supervisor editable no (locked)
Stage Parameter Factory setting
Degassing Duration 1 minutes
Washing
Chemical dose 0 ml (0.0 fl.oz)
Temperature +40 °C (+104 °F)
Duration 30 minutes
Rinse
Temperature 0 °C (+32 °F)
1
Duration 1 minute
1
A setting of 0 °C (+32 °F) is interpreted at 'temperature not important' by the equipment. The
tank will fill with water from the supply hoses, the equipment will not attempt to regulate the
incoming water temperature, or adjust the temperature of the water in the tank.
106 WASH PROGRAMS
InnoWave Unity & Unity 20 Operator Manual