5.3 Loading General Instruments
Instruments with no internal surface (non- cannulated), such as clamps, scissors, and
forceps, can be placed directly into the instrument basket.
1. If not already removed, remove the basket from the tank, and move to a loading
area away from the sonic irrigator.
2. Open mechanical instruments, such as scissors and clamps.
34 USING THE INSTRUMENT BASKET
InnoWave Unity & Unity 20 Operator Manual