Page #178 background image
Commands listed in alphabetical order
Arguments
FALLING decode
s on the falling edge of the clocked parallel bus signal.
RISING decodes on the rising edge of the clocked parallel bus signal.
EITHER decodes on the rising or falling edge of the clocked parallel bus signal.
Examples
BUS:B1:PARALLEL:CLOCK:EDG E FALLING sets the decoding to happen on the
falling edge of its clocked parallel bus B1.
BUS:B1:PARALLEL:CLOCK:EDG E? might return
:BUS:B1:PARALLEL:CLOCK:ED GE RISING, indicating that when bus B1
operates in Clocked Parallel mode, it samples new data values on the rising edge
of its clock source signal.
BUS:B<x>:PARallel:CLOCk:ISCLOCKED
This command determines whether the bus opera tes in a clocked or async hronous
fashion. The bus is specied by x.
Group
Bus
Syntax
BUS:B<x>:PARallel:CLOCk:I SCLOCKED {OFF|ON|NR1>
BUS:B<x>:PARallel:CLOCk:I SCLOCKED?
Arguments
OFF argument species an asynchronous bus.
ON argument species a clocked bus.
<NR1> =0species an asynchronous bus; a ny other value species a clocked bus.
Examples
BUS:B1:PARallel:CLOCk:ISC LOCKED 0 sets the bus to operate
asynchronously.
BUS:B1:PARallel:CLOCk:ISC LOCKED? might return
:BUS:B1:PARALLEL:CLOCK:IS CLOCKED 1 indicating the bus is
clocked.
BUS:B<x>:PARallel:CLOCkSOUrce
This command sets or queries the Parallel c lock bit source for the specied bus.
The bus is specied by x.
Group
Bus
2-162 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Other manuals for Tektronix MSO54

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54 and is the answer not in the manual?

Tektronix MSO54 Specifications

General IconGeneral
BrandTektronix
ModelMSO54
CategoryTest Equipment
LanguageEnglish

Related product manuals