Commands listed in alphabetical order
SEARCH:SEARCH
1:TRIGGER:A:LOGIC:FUNCTION?
might return
:SEARCH:SEARCH1:TRIGGER:A: LOGIC:FUNCTION NOR, indicating that the
logic function for search 1 is set to NOR.
SEARCH:SEAR
CH<x>:TRIGger:A:LOGIc:INPUT:CLOCK:SOUrce
This command species or queries the channel to use as the clock source for logic
trigger. The search number is specied by <x>.
Group
Search and Mark
Syntax
SEARCH:SEARCH<x>:TRIGger:A :LOGIc:INPUT:CLOCK:SOUrce
{CH<x>|Ch<x>_D<x>|REF<x>_D <x>}
SEARCH:SEARCH<x>:TRIGger:A :LOGIc:INPUT:CLOCK:SOUrce?
Arguments
Arguments are the possible input channels.
Examples
SEARCH:SEARCH1:TRIGger:A:L OGIc:INPUT:CLOCK:SOUrce C H2 sets the
clock source to channel 2.
SEARCH:SEARCH1:TRIGger:A:L OGIc:INPUT:CLOCK:SOUrce? might return
:SEARCH:SEARCH1:TRIGGER:A: LOGIC:INPUT:CLOCK:SOURCE UNDEFINED
indicating the clock source is not dened.
SEARCH:SEARCH<x>:TRIGger:A:LOGIc:LEVel:CH<x>
This command sets or queries the voltage level to use for logic trigger search.
Thesearchnumberisspecied by <x>.
Group
Search and Mark
Syntax
SEARCH:SEARCH<x>:TRIGger:A :LOGIc:LEVel:CH<x> <NR3>
SEARCH:SEARCH<x>:TRIGger:A :LOGIc:LEVel:CH<x>?
Arguments
<NR3> is the voltage level to use for logic trigger search.
Examples
SEARCH:SEARCH1:TRIGger:A:L OGIc:LEVel:CH2 50.0e-3 sets the level
to 50 mV.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-717

Table of Contents

Other manuals for Tektronix MSO54

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54 and is the answer not in the manual?

Tektronix MSO54 Specifications

General IconGeneral
BrandTektronix
ModelMSO54
CategoryTest Equipment
LanguageEnglish

Related product manuals