Commands listed in alphabetical order
Group
Vertical
Syntax
CH<x>:PRObe:ID:TYPe?
Examples
CH1:PROBE:ID:TYPE? might return :CH1:PROBE:I D:TYPE "TCP0030",
indicating that a TCP0030 current probe is attached to Channel 1.
CH<x>:PRObe:RESistance? (Query Only)
This query-only command returns the resistance of the probe that is attached to
the specied channel. The channel is specied by x.
Group
Vertical
Syntax
CH<x>:PRObe:RESistance?
Examples
CH2:PROBE:
RESISTANCE?
might return :CH2:PROBE:RES ISTANCE
1.0000E+06
, indicating that the input resistance of the probe attached to
Channel 2 is 1 M.
CH<x>:PRObe:SET
This command sets or queries a spects of probe accessory user interfaces, for
example probe attenuation factors or probe audible over range. The available
arguments for this command will vary depending on the accessory you attach to
the instrument. The channel is specied by x.
Group
Vertical
Syntax
CH<x>:PRObe:SET <QString>
CH<x>:PRObe:SET?
Arguments
<QString> is a quoted string representing a settable aspect of the attached
accessory.
Examples
CH6:PRObe:SET "ATTENUATION 5X" setting the probe to 25X attenuation.
CH6:PRObe:SET? might return :CH6:PROBE: SET "ATTENUATION
25X"
indicating that the probe is set to the selection for 25X attenuation.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-199

Table of Contents

Other manuals for Tektronix MSO54

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54 and is the answer not in the manual?

Tektronix MSO54 Specifications

General IconGeneral
BrandTektronix
ModelMSO54
CategoryTest Equipment
LanguageEnglish

Related product manuals