Commands listed in alphabetical order
Syntax
MEASUrement:C
H<x>:REFLevels:ABSolute:HYST eresis <NR3>
MEASUrement:CH<x>:REFLeve ls:ABSolute:HYSTeresis?
Arguments
<NR3> is the hysteresis value us ed for autoset.
Examples
MEASUREMENT:CH2:REFLEVELS :ABSOLUTE:HYSTERESIS 30.0 000E-3 sets
the reference hysteresis level to 30 mV.
:MEASUREMENT:CH2:REFLEVEL S:ABSOLUTE:HYSTERESIS? might return
:MEASUREMENT:CH2:REFLEVEL S:ABSOLUTE:HYSTERESIS 30. 0000E-3
indicating that reference hysteresis level is set to 30 mV.
MEASUrement:CH<x>:REFLevels:ABSolute:RISEHigh
This command sets or queries the value used as the high reference level of the
rising edge when the source ref level method is set to absolute. The channel
number is specified by x.
Group
Measurement
Syntax
MEASUrement:CH<x>:REFLeve ls:ABSolute:RISEHigh <NR3 >
MEASUrement:CH<x>:REFLeve ls:ABSolute:RISEHigh?
Arguments
<NR3> is the high r eference level of the rising edge when the source ref level
method is set to absolute.
Examples
MEASUREMENT:CH2:REFLEVELS :ABSOLUTE:RISEHIGH 1.5 sets the h igh
reference level of the rising edge to 1.5 V.
MEASUREMENT:CH2:REFLEVELS :ABSOLUTE:RISEHIGH? might return
:MEASUREMENT:CH2:REFLEVEL S:ABSOLUTE:RISEHIGH 1.000 0 indicating
that the high reference level of the rising edge is 1.0 V.
MEASUrement:CH<x>:REFLevels:ABSolute:RISELow
This command sets or queries the value used as the low reference level of the
rising edge when the source ref level method is set to absolute. The channel
number is specified by x.
Group
Measurement
2-384 MSO54, MSO56, MSO58, MSO58LP Programmer