Commands listed in alphabetical order
NOCARE disable
s filtering of command words on the parity bit.
Examples
TRIGGER:A:BUS:B1:MIL1553B: COMMAND:PARITY ONE specifies filtering
command words for thos
ewheretheparitybithasavalueof1.
TRIGGER:A:BUS:B1:MIL1553B: COMMAND:PARITY? might return NOCARE,
indicating that comm
and words are not being filtered based on the parity bit value.
TRIGger:{A|B}:BUS:B<x>:MIL1553B:COMMAND:SUBADdress
This command sets or queries the value of the command word subaddress field for
a MIL-STD-1553 bus to triggering on. The bus number is specified by x. The
trigger condition must be set to COMMAND.
Conditions
Requires the SR-AERO Triggering and Analysis application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:MIL 1553B:COMMAND:SUBADdress <QString>
TRIGger:{A|B}:BUS:B<x>:MIL 1553B:COMMAND:SUBADdress?
Arguments
<QString> is the subaddress value.
Examples
TRIGGER:A:BUS:B1:MIL1553B: COMMAND:SUBADDRESS "X1000 " sets the
value to X1000.
TRIGGER:A:BUS:B1:MIL1553B: COMMAND:SUBADDRESS? might return
"XXXXX", indicating that the value is XXXXX.
TRIGger:{A|B}:BUS:B<x>:MIL1553B:COMMAND:TRBit
This command sets or queries the value of the command word Transmit / Receive
bit for a MIL-STD-1553 bus to trigger on. The bus number is specified by x
.The
trigger condition must be set to COMMAND.
Conditions
Requires the SR-AERO Triggering and Analysis application.
Group
Trigger
MSO54, MSO56, MSO58, MSO58LP Programmer 2-817