Commands listed in alphabetical order
Arguments
FORWard speci
es a forward search to the edge.
BACKWard speci es a backward search to the edge.
Examples
MEASUrement:MEAS1:TOEDGES EARCHDIRect FORWARD species a forward
search to the edge.
MEASUrement:MEAS1:TOEDGES EARCHDIRect? might return
:MEASUREMENT:MEAS1:TOEDGE SEARCHDIRECT FORWARD indicating the
instrument will search in the forward direction for the to edge.
MEASUrement:MEAS<x>:TRANSition
This command sets or queries the transition edges ag for the meas urement. The
measurement number is specied by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:TRANS ition {<NR1>|OFF|ON}
Arguments
<NR1> = 1, the measurement is computed on rising (if measurement type is
rise time) or falling edges (if measurement type is fall time) following a double
transition only. If it is set to 0 , the measurement is computed on all rising (if
measurement type is rise time) or falling (if measurement type is fall time) edges.
OFF compute s the measurement on all rising (if measurement type is rise time)
or falling (if measurement type is fall time) edges.
ON computes the measurement on rising (if measurement type is rise time) or
falling edges (if measurement type is fall time) following a double transition only.
Examples
MEASUrement:MEAS1:TRANSit ion 1 specifyingthe measurement is computed
on rising (if measurement type is rise time) or falling edges (if measurement type
is fall time) following a double transition only.
MEASUrement:MEAS1:TRANSit ion? might return
:MEASUREMENT:MEAS1:TRANSI TION 0 indicating the measurement is
computed on all rising (if measurement type is rise time) or falling (if
measurement type is fall time) edges.
MEASUrement:MEAS<x>:TYPe
This command sets or queries the measurement type. The measurement number
is specied by x.
2-488 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Other manuals for Tektronix MSO54

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54 and is the answer not in the manual?

Tektronix MSO54 Specifications

General IconGeneral
BrandTektronix
ModelMSO54
CategoryTest Equipment
LanguageEnglish

Related product manuals