EasyManuals Logo
Home>Tektronix>Test Equipment>MSO54

Tektronix MSO54 Programmer's Manual

Tektronix MSO54
1026 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #462 background imageLoading...
Page #462 background image
Commands listed in alphabetical order
MEASUrement:
MEAS<x>:EDGES:TOLevel
This command sets or queries the 'to level' edge for the measurement.
Measurements are specied by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:EDGES :TOLevel {HIGH|MID|LOW}
MEASUrement:MEAS<x>:EDGES :TOLevel?
Arguments
HIGH species the H IGH level.
MID species the MID level.
LOW species the LOW level.
Examples
MEASUrement:MEAS1:EDGES:T OLevel HIGH species the High to level.
MEASUrement:MEAS1:EDGES:T OLevel? might return
:MEASUREMENT:MEAS1:EDGES: TOLEVEL LOW indicating the to level
edge is set to the Low level.
MEASUrement:MEAS<x>:EDGES:UPPERFREQuency
This command sets o r queries the upper frequency for the measurement when the
measurement type is phase noise. Higher frequencies are ignored. Measurements
are specied by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:EDGES :UPPERFREQuency <NR3>
MEASUrement:MEAS<x>:EDGES :UPPERFREQuency?
Arguments
<NR3> is the upper frequency of the edge.
Examples
MEASUrement:MEAS1:EDGES:U PPERFREQuency 1.5 sets the upper frequency
to 1.5 MHz.
MEASUrement:MEAS1:EDGES:U PPERFREQuency? might return
:MEASUREMENT:MEAS1:EDGES: UPPERFREQUENCY 1.0000E+6 indicating the
upper frequency is 1.0 MHz.
2-446 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Other manuals for Tektronix MSO54

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54 and is the answer not in the manual?

Tektronix MSO54 Specifications

General IconGeneral
BrandTektronix
ModelMSO54
CategoryTest Equipment
LanguageEnglish

Related product manuals