Warnings cautions and notices 11
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
2 Warnings cautions and notices
2.1 Always observe
NOTE: The use of the ventilator requires detailed knowledge and the
understanding of this operator’s manual. This device is only intended for the
described use.
WARNING: Only use this ventilator in combination with an external
monitoring device (for example: SpO
2
).
WARNING: Only operate the ventilator with accessories recommended by
ACUTRONIC Medical Systems AG.
WARNING: The ventilator must be operated by qualified clinical staff to
ensure immediate remedial action in the event of malfunction.
WARNING: The fabian system and associated auxiliary systems must
NEVER be used in MRI scanning facilities.
WARNING: An alternate ventilation method (for example manual
resuscitation) must always be available when using the ventilator.
WARNING: DO NOT use the ventilator in combination with flammable gases
or narcotic agents or in an oxygen-rich environment to prevent the risk of fire
or explosion.
WARNING: NEVER use the ventilator in explosive environments.
WARNING: An audible signal indicates a system or patient alarm and always
requires action by a trained medical professional.
WARNING: If an alarm condition (other than the exceptions listed within this
manual) occurs while the audible alarm Silence function is engaged, only the
visual alarm indications are displayed.
WARNING: DO NOT silence an audible alarm, engage the audible Alarm
Silence function, or decrease the audible alarm volume if patient safety could
be compromised.
WARNING: DO NOT obstruct the speaker.
Blocking the speaker can result in an inaudible alarm tone.
WARNING: Carefully route patient cabling to reduce the risk of patient
entanglement or strangulation.
WARNING: NEVER connect the ventilator to a patient if an error or
malfunction is detected during equipment check.
WARNING: NEVER connect to electrical devices not mentioned in this
operator’s manual without first consulting the manufacturer.
WARNING: Connect only electrical devices which are IEC60601-1 approved.
WARNING: NEVER operate the ventilator while covered or set up in a way
to negatively impact the operation or function.