EasyManuals Logo
Home>ACUTRONIC>Medical Equipment>fabian HFO

ACUTRONIC fabian HFO User Manual

ACUTRONIC fabian HFO
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #13 background imageLoading...
Page #13 background image
Warnings cautions and notices 13
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
#
Symbol
Description
34.
WARNING: Check alarm limit settings each time the etCO
2
module is used.
35.
WARNING: The etCO
2
module is intended only as an adjunct in patient
assessment. It must be used in conjunction with assessment of clinical signs
and symptoms.
36.
Before use, carefully read the Oximetry Sensor Directions for Use and PC-
Series Patient Cable Directions for Use.
37.
EXPLOSION HAZARD: Do not use the pulse oximeter in the presence of
flammable anesthetics or other flammable substances in combination with
air, oxygen-enriched environments, or nitrous oxide.
38.
Check alarm limit settings each time the pulse oximeter is used.
39.
A pulse oximeter should NOT be used as an apnea monitor.
40.
A pulse oximeter should be considered an early warning device. As a trend
towards patient deoxygenation is indicated, blood samples should be
analyzed by a laboratory co-oximeter to completely understand the patient's
condition.
41.
Always remove the sensor from the patient and completely disconnect the
patient from the pulse oximeter before bathing the patient.
42.
Do not use malfunctioning equipment. Have the pulse oximeter unit repaired
by Masimo or a qualified service person.
43.
ELECTRIC SHOCK HAZARD: Do not remove the pulse oximeter cover.
There are no user-serviceable items inside the oximeter. An operator may
only perform maintenance procedures specifically described in this manual.
ONLY connect IEC 60601-1 or IEC 60950-1 compliant devices to Ethernet,
Nurse call, and RS232 ports.
44.
If the accuracy of any measurement by the pulse oximeter does not seem
reasonable, first check the patient's vital signs by alternate means, and then
check the oximeter for proper functioning.
45.
WARNING: Do not use the ventilator in association with HF (High
Frequency) electrosurgical equipment.
46.
WARNING: Connect SpO
2
and etCO
2
sensor cables to the machine before
the patient is connected.
2.2 Maintenance
The device is a ventilator classified as Class IIb according to the European Directive, as such:
Inspection according to manufacturer specifications is required every 12 months.
Maintenance must be performed by ACUTRONIC Medical Systems trained personnel with
access to appropriate test and measuring equipment.
We strongly recommend ACUTRONIC Medical Systems AG exclusive representatives for service
agreements and repairs.
Only use original ACUTRONIC Medical Systems parts for repairs.
Note chapter ”Service and maintenance intervals”.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian HFO and is the answer not in the manual?

ACUTRONIC fabian HFO Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian HFO
CategoryMedical Equipment
LanguageEnglish

Related product manuals