82 Alarms
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
#
Alarm Text
Alarm Type
Cause
Solution
Main
Priority
Sub
Priority
64.
High minute
volume
Patient alarms
Lung compliance
has increased.
Resistance has
decreased.
Hyperventilation.
Check patient,
Check ventilation
settings.
Adjust settings
II
8
65.
Low minute
volume
Patient alarms
Lung compliance
has decreased.
Resistance has
increased.
Spontaneous
breathing has
stopped or is
declining
Check patient,
Check tubing
system for
obstructions,
Check ventilation
settings
II
8
66.
Low PEEP
Patient alarms
Leak or
disconnection.
Insp. flow or exp.
flow set too low.
Check patient,
Check tubing
system for tight
connection.
Increase flow.
II
8
67.
VTe not
reached /
check settings
System limits
Inspiratory pressure
default is reached
prematurely.
Patient compliance has
worsened
Check patient,
Adjust inspiratory
pressure setting.
Increase inspiratory
period.
Adjust volume
setting
II
8
68.
Inspiratory
pressure not
reached
System limits
Leak.
Flow too low.
Check patient,
Check tubing
system for tight
connection.
Increase flow.
III
9
69.
Power failure
Electrical power
Power failure/
Ventilator was
unplugged from mains
power supply and is
operating on battery
power.
Switch to mains
supply.
Check power cord
for loose
connection.
Alarm can be
silenced with Alarm
Slience button.
The device
continues in battery
mode without
interruption.
III
9
70.
Tidal Volume
limited
System limits
Upper alarm limit
exceeded
Adjust settings.
III
9
7.4.2 Pressure release behaviour

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian HFO and is the answer not in the manual?

ACUTRONIC fabian HFO Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian HFO
CategoryMedical Equipment
LanguageEnglish

Related product manuals