EasyManuals Logo
Home>ACUTRONIC>Medical Equipment>fabian HFO

ACUTRONIC fabian HFO User Manual

ACUTRONIC fabian HFO
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #83 background imageLoading...
Page #83 background image
Alarms 83
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
In case of detected obstruction of the breathing tubes, the ventilator alarms and also ceases gas flow
to the outlet, and opens the exhalation valve to reduce the pressure in the breathing circuit to
ambient pressure (ZEEP). The tube obstruction alarm is terminated once the airway pressure
measured by the proximal pressure sensor has fallen below set PEEP and the time of the next
breath trigger is reached.
This pressure relief occurs:
12 mbar above the set PIP in IPPV, SIPPV, SIMV, SIMV+PSV, PSV and CPAP modes.
10 mbar above the set PIP in NCPAP and DUOPAP modes.
In case of detected high inspiratory pressure, the ventilator alarms and switches to the expiratory
phase the exhalation valve pressure control is reduced to PEEP, and the outlet gas flow is
switched from I-Flow to E-Flow. Only if this subsequently transitions to a tube obstruction alarm, the
outlet flow is switched off and exhalation valve opened as above.
7.4.3 Application error
If the device software encounters an unexpected ERROR, a window appears in the middle of the
screen with the following text in English: “Application fabianHFO.exe encountered a serious Error
and must shut down.”
If there is an audible alarm, follow these steps:
1. Immediately start ventilation of patient with an Alternative Ventilation method
2. Shut down the device.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian HFO and is the answer not in the manual?

ACUTRONIC fabian HFO Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian HFO
CategoryMedical Equipment
LanguageEnglish

Related product manuals