132 Accessories and options
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
12.6.3 Erasing the FOT graph
The FOT Graph is cleared if any of the following actions are performed:
a new FOT sequence is started,
the device is restarted,
the patient range is changed,
device date / time is changed
12.6.4 FOT disabling conditions
FOT is turned OFF and cannot be re-enabled while any of the following actions occurs or alarms are
active:
Any system alarm:
Air supply pressure
Check Blender
Exhalation calibration
Oxygen supply pressure
Tube Occlusion
Any flow sensor alarm:
Clean flow sensor
Flow sensor defect
Flow sensor not connected
Disconnection / Tube blocked alarm:
Check ET tube
Patient disconnected
Any system failure:
Checksum conductor PIC
Checksum monitor PIC
COM interface
Cooling fan defect
DIO interface
I2C interface
Input pressure blender
Low physical memory
parallel interface
Safety relay defect
SPI interface
Voltage monitoring
Alarm actions:
Applying Manual breath
Applying or preset another ventilation mode

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian HFO and is the answer not in the manual?

ACUTRONIC fabian HFO Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian HFO
CategoryMedical Equipment
LanguageEnglish

Related product manuals