132 Accessories and options
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
12.6.3 Erasing the FOT graph
The FOT Graph is cleared if any of the following actions are performed:
• a new FOT sequence is started,
• the device is restarted,
• the patient range is changed,
• device date / time is changed
12.6.4 FOT disabling conditions
FOT is turned OFF and cannot be re-enabled while any of the following actions occurs or alarms are
active:
Any system alarm:
• Air supply pressure
• Check Blender
• Exhalation calibration
• Oxygen supply pressure
• Tube Occlusion
Any flow sensor alarm:
• Clean flow sensor
• Flow sensor defect
• Flow sensor not connected
Disconnection / Tube blocked alarm:
• Check ET tube
• Patient disconnected
Any system failure:
• Checksum conductor PIC
• Checksum monitor PIC
• COM interface
• Cooling fan defect
• DIO interface
• I2C interface
• Input pressure blender
• Low physical memory
• parallel interface
• Safety relay defect
• SPI interface
• Voltage monitoring
Alarm actions:
• Applying Manual breath
• Applying or preset another ventilation mode