162 Technical specifications
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
18.12 Essential Performance
#
Essential Performance Function
Standard
Clause
1
Delivery of ventilation at the patient-
connection port within the alarm limits set
by the operator;
1
or
the following alarm conditions shall occur
as appropriate:
Low battery as internal electrical
power source nears depletion;
Low and High Minute Volume;
Low and High PEEP;
Obstruction/Occlusion;
Low and High Air/Oxygen Inlet
Pressure;
Air/Oxygen Gas Supply Failure;
or
a technical alarm condition shall occur.
ISO 80601-2-12:
2011
202.6.2.1.10
201.11.8.101.2
201.12.4.103
201.12.4.106
201.12.4.107
201.11.8.101.1
201.11.8.101.2
201.13.101/.102
201.101.1
2
Measurement accuracy of the oxygen
respiratory gas monitor (RGM) is
maintained within the IFU specification, ±
(3% absolute);
or
Low and High Oxygen Level alarm
conditions shall occur, as appropriate;
or
a technical alarm condition shall occur.
ISO 80601-2-12:
2011
201.12.4.101
ISO 80601-2-55:
2018
201.12.1.101.1
208.6.1.2
201.11.8.101.1
3
Measurement accuracy of the etCO
2
RGM is
maintained within the IFU specification:
0 to 38 mmHg
± 2 mmHg
39 to 150 mmHg
± (5% of reading + 8% of (reading 39
mmHg));
or
Low and High CO
2
Level alarm conditions
shall occur, as appropriate;
or
a technical alarm condition shall occur.
ISO 80601-2-55:
2018
201.12.1.101.1
208.6.1.2
201.11.8.101.1
4
Measurement accuracy of the SpO
2
monitor
is maintained within the IFU specification
only within the range 70 to 100%:
± 2% pediatric patients, no motion
conditions;
± 3% pediatric patients, motion
conditions, and all neonatal patients;
or
Low and High SpO
2
Level alarm conditions
shall occur, as appropriate;
or
a technical alarm condition shall occur.
ISO 80601-2-61:
2017
201.12.1.101.1
208.6.1.2.101
201.11.8.101.1
201.12.4
201.13.101
5
Measurement accuracy of the pulse rate
from the SpO
2
monitor is maintained within
ISO 80601-2-61:
2017
201.12.1.104

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian HFO and is the answer not in the manual?

ACUTRONIC fabian HFO Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian HFO
CategoryMedical Equipment
LanguageEnglish

Related product manuals