Do not use any accessory if individual package is damaged.
If the packaging of the ventilator is damaged use the device only if the startup
device check is performed successfully.
For initial installation and when lifting the fabian HFO, disconnect all cables and circuits from the
ventilator, grab the ventilator with both hands from each lateral side, see arrows indicating on the left
side in the figure below, and lift the ventilator carefully.
Always secure the fabian HFO using the dedicated screw at the bottom of the device.
5.1.1 Connect the power supply
Connect the fabian with power cable to a suitable power outlet.
WARNING:
DO NOT connect the device to a power outlet strip.
Power fluctuations from the system may trip the circuit breaker in the power outlet strip
causing loss of power to the device.
Exception: An approved medical grade power outlet strip with valid amperes rating by the
manufacturer for connection to a ventilator.
The device can be operated with 100 to 240 VAC and 50 or 60 Hz and automatically adjusts to
voltage and frequency without manual switch-over being necessary.
WARNING:
The use of fuses must correspond with the value and type approved for the fabian.
• The fuse type and value is printed on the back of the fabian.
• Replacement of the fuses can only be carried out by trained staff. Replace fuses with
identical type and values. Failing to do so, can cause fire hazards.