160 Technical specifications
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
18.8 Data storage
Alarms
Maximum: 1.000 messages.
Log
Maximum: 10 Log Files.
Log is stored during power failure.
When Log capacity is reached index, is shifted and oldest log file deleted.
Trend
Maximum: 5 days.
Storage: every 30 seconds.
18.9 Applied parts
Applied parts for the device are Nebulizer, Flow Sensor, SpO2 module (type BF) and tube set
(type B).
18.10 Internal device checks
The following internal device checks are performed on Startup:
COM: Communication between GUI and Conductor PIC.
DIO: Communication between GUI and Digital I/O.
I
2
C: Communication between GUI to Accu PIC, NVRAM and Multiplexer.
PIF: Communication between GUI and Parallel Interface
Relay: Check from the SPI
SPI: Communication between GUI and Monitor PIC.
SPICks: SPI Checksum
Voltage: Voltage Error Check from the SPI
18.11 Gas blender function
The Air / Oxygen Blender provides a Gas blend with adjustable proportions of Oxygen and Air.
Gas from the central gas supply enters the device through the gas input connections.
Blending of the gases occurs through two proportional valves.
The proportional valves, which also act as return valves, prevent the return flow of a gas into the
supply line of the other gas.An Oxygen sensor measures the Inspiratory Oxygen Concentration.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian HFO and is the answer not in the manual?

ACUTRONIC fabian HFO Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian HFO
CategoryMedical Equipment
LanguageEnglish

Related product manuals