To Secure Against Calibration You can secure the function
generator against accidental calibration either from the front panel or
over the remote interface. The function generator is secured when
shipped from the factory and the security code is set to “HP033120”.
Be sure to read the security code rules on page 118 before attempting to
secure the function generator.
Front-Panel Operation:
1: UNSECURED (CAL MENU)
If the function generator is unsecured, you will see the above
command when you go into the
CAL MENU. To secure the function
generator, go to the “
PARAMETER level of the UNSECURED
command, enter the security code, and then press Enter.
When you go to the “
COMMANDS level in the CAL MENU again,
you will notice that the function generator is secured. Notice also that
the “
CALIBRATE command is now hidden and you cannot perform a
calibration.
Remote Interface Operation:
CALibration:SECure:STATe {OFF|ON},<
code>
To secure the function generator, send the above command with the
same code as used to unsecure calibration. For example,
"CAL:SEC:STAT ON,HP033120"
¾000000:CODE
1: SECURED
Chapter 3 Features and Functions
Calibration Overview
120

Table of Contents

Other manuals for Agilent Technologies 33120A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33120A and is the answer not in the manual?

Agilent Technologies 33120A Specifications

General IconGeneral
BrandAgilent Technologies
Model33120A
CategoryInverter
LanguageEnglish

Related product manuals