To Unsecure for Calibration You can unsecure the function
generator for calibration either from the front panel or over the remote
interface. The function generator is secured when shipped from the
factory and the security code is set to “HP033120”.
Front-Panel Operation:
1: SECURED (CAL MENU)
If the function generator is secured, you will see the above command
when you go into the
CAL MENU. (If you move across the “COMMANDS
level in the menu, you will notice that the “
CALIBRATE command is
hidden when the function generator is secured.) To unsecure the
function generator, go to the “
PARAMETER level of the SECURED
command, enter the security code, and then press Enter.
When you go to the “
COMMANDS level in the CAL MENU again,
you will notice that the function generator is unsecured. Notice also
that the “
CALIBRATE command is no longer hidden and you can
perform a calibration.
Remote Interface Operation:
CALibration:SECure:STATe {OFF|ON},<
code>
To unsecure the function generator, send the above command with
the same code used to secure calibration. For example,
"CAL:SEC:STAT OFF,HP033120"
¾000000:CODE
1: UNSECURED
3
Chapter 3 Features and Functions
Calibration Overview
119

Table of Contents

Other manuals for Agilent Technologies 33120A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33120A and is the answer not in the manual?

Agilent Technologies 33120A Specifications

General IconGeneral
BrandAgilent Technologies
Model33120A
CategoryInverter
LanguageEnglish

Related product manuals