Page #66 background image
Output Units
Applies only to output amplitude (does not affect offset). At power-on,
the units for output amplitude are volts peak-to-peak.
Output units: Vpp, Vrms, or dBm. The default is Vpp.
The unit setting is stored in volatile memory; the units are set to
Vpp” when power has been off or after a remote interface reset.
When switching from remote programming over the interface back to
front-panel (local) operation, the output units are remembered.
For example, if you select “Vrms from the remote interface, the units
are shown as “Vrms” on the front-panel display.
Front-Panel Operation: Use theEnter Number” units keys
(the arrow keys located on the right side of the front panel) to select
the output units. You can also convert from one unit to another from
the front panel. For example, to convert 2 Vpp to its equivalent value
in Vrms, press the following keys:
The converted value is 707.1 mVrms.
Remote Interface Operation:
VOLTage:UNIT {VPP|VRMS|DBM|DEFault}
kHz
m Vrms
¿ Enter Number
Chapter 3 Features and Functions
Output Configuration
64

Table of Contents

Other manuals for Agilent Technologies 33120A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33120A and is the answer not in the manual?

Agilent Technologies 33120A Specifications

General IconGeneral
BrandAgilent Technologies
Model33120A
CategoryInverter
LanguageEnglish

Related product manuals