Burst Source
In the external gated burst mode, the output waveform is either
on or “off based on the level of the external signal applied to the
rear-panel Ext Trig terminal. When the gate signal is true, the function
generator outputs a continuous waveform. When the gate signal is false,
the output is turned off.
Burst source: Internal or External Gate. The default is Internal.
When the internal burst source is selected, the external gated mode
is disabled.
When the external gate source is selected, the output is enabled or
disabled based on the logic level of the gate signal applied to the
Ext Trig terminal. When the gate signal is true (
TTL high),
the function generator outputs a continuous waveform. When the
gate signal is false (
TTL low), the output is turned off (zero volts or
the dc offset level).
When the external gate source is selected, the burst count, burst rate,
burst phase, and burst trigger source are ignored (these parameters
are used for the triggered burst mode only).
The Ext annunciator turns on when the function generator is waiting
for the external gate signal.
Front-Panel Operation:
7: BURST SRC (MOD MENU)
Remote Interface Operation:
BM:SOURce {INTernal|EXTernal}
3
Chapter 3 Features and Functions
Burst Modulation
83

Table of Contents

Other manuals for Agilent Technologies 33120A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33120A and is the answer not in the manual?

Agilent Technologies 33120A Specifications

General IconGeneral
BrandAgilent Technologies
Model33120A
CategoryInverter
LanguageEnglish

Related product manuals