-151 Invalid string data
An invalid character string was received. Check to see if you have
enclosed the character string in single or double quotes and that the
string contains valid
ASCII characters.
-158 String data not allowed
A character string was received but is not allowed for the command.
Check the list of parameters to verify that you have used a valid
parameter type. Example: BMOD:NCYC ’TEN’
-161 Invalid block data
Applies only to the DATA:DAC VOLATILE command. For a definite-
length block, the number of bytes of data sent does not match the
number of bytes that you specified in the block header. For an
indefinite-length block, an EOI (End-or-Identify) was received without
an accompanying <new line> character.
-168 Block data not allowed
Block data was received but is not allowed for the command. Check to
see if you have sent the correct data type with the command.
-170 to -178 Expression errors
The function generator does not accept mathematical expressions.
-211 Trigger ignored
A Group Execute Trigger (GET) or
*TRG was received but the trigger
was ignored. Make sure that you have selected the correct trigger source.
5
Chapter 5 Error Messages
Execution Errors
231

Table of Contents

Other manuals for Agilent Technologies 33120A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 33120A and is the answer not in the manual?

Agilent Technologies 33120A Specifications

General IconGeneral
BrandAgilent Technologies
Model33120A
CategoryInverter
LanguageEnglish

Related product manuals