7-2 ApexPro™ 2001989-351A
Checkout
Antenna checkout
Procedure
1. If the antenna is active, verify that the power LED on the antenna is lit.
2. Perform visual inspection. See Visual inspection on page 4-4.
3. For the replaced antenna, perform the Unity Signal Gain Test with Transmitter
from the Antenna system installation verification tests section of the ApexPro
CH Antenna System Test Instructions.
4. For the field that the replaced antenna is on, perform the Out-of-Band RF Signal
Test from the Antenna system installation verification tests section of the
ApexPro CH Antenna System Test Instructions.
Infrastructure equipment checkout
This section applies to: power supplies, power cords, bias tees, antenna amplifiers,
splitters/combiners, attenuators, filters, coaxial cables, connectors, and adapters.
Procedure
1. If replacing coaxial cables or connectors, perform a distance-to-fault or other
similar test to verify continuity of the cable.
2. Perform visual inspection. See Visual inspection on page 4-4.
3. Multiple antennas may be connected to the replaced component. For each
affected antenna, perform the Unity Signal Gain Test with Transmitter from the
Antenna system installation verification tests section of the ApexPro CH
Antenna System Test Instructions. If there are more than 10 affected antennas, it
is sufficient to perform the test on only 10 of the affected antennas.
4. For the field that the replaced component is on, perform the Noise Floor
Performance Test from the Antenna system installation verification tests section
of the ApexPro CH Antenna System Test Instructions.
5. For the field that the replaced component is on, perform the Out-of-Band RF
Signal Test from the Antenna system installation verification tests section of the
ApexPro CH Antenna System Test Instructions.
Receiver subsystem checkout
Checkout procedure
1. Perform visual inspection. See Visual inspection on page 4-4.
2. Perform receiver system checkout. See Receiver system checkout on page 7-3.
3. Perform electrical safety tests. See Electrical safety tests on page 7-7.

Table of Contents

Other manuals for GE ApexPro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE ApexPro and is the answer not in the manual?

GE ApexPro Specifications

General IconGeneral
BrandGE
ModelApexPro
CategoryMedical Equipment
LanguageEnglish

Related product manuals