A-6 ApexPro™ 2001989-351A
Technical specifications
ECG
Environmental specifications
Operating conditions
Transport and storage conditions
Device specifications
Multi-channel configuration
(5- or 6- leadwire)
I, II, III, Va, Vb, aVR, aVL, aVF
Leads analyzed
simultaneously
Four (I, II, III, V)
Single-channel (3-leadwire)
configuration
I, II or III, configurable
Heart rate detection 30 to 300 BPM
QRS detection range 0.5 to 5 mV
Frequency response up to 38Hz (-3 dB)
A/D converter resolution 10 bits, 9.75 µV (RTI)
Sample rate 240 samples/seconds
Temperature 5 to 40° C
Relative humidity 15 to 95% (non-condensing)
Temperature -40 to 70° C
Relative humidity 15 to 95% (non-condensing)
Pressure 475 to 1200 hPa
Water resistance IEC 60529 IPX7 rating (can survive inadvertent submersion)
Input configuration 3, 5 or 6 electrodes
Frequency range 1395.025 - 1399.975 MHz
Power output: 5mW
(Meets FCC requirements for Part 95.)
Bandwidth 11.9 KHz
Modulation GFSK
Serial I/O ports 2
Alarm pause Transmitted and indicated via LED