Equipment Overview
2001989-351A ApexPro™ 2-9
Transmitter controls, indicators and labels
Views
The transmitters have the following buttons and LEDs:
605A, 207A, 206B
ApexPro, ApexPro CH, and T14 Transmitters
ApexPro ApexPro CH & T14 Function
RL RA LA LL Va Vb RL RA LA LL Va Vb
N R L F Ca Cb
When first powered up, the lead LEDs flash
rapidly, followed by two slow flashes. The
transmitter begins functioning after the two
slow flashes.
When any of the transmitter's buttons are
pushed, the lead LEDs flash twice.
Change Battery When the battery power is running low, the
change battery LED flashes.
Verify Leads
When pressed, the lead LEDs flash twice. If a
lead is valid, its LED stays lit for one minute.
Pause Alarm
When the Pause Alarm condition occurs, the
pause alarm LED flashes until the condition
ends.